URIT 3000PLus Hematology Analyzer

Parameters:

WBC, LYM#, MID#, GRAN#, LYM%, MID%, GRAN%, RBC, HGB, HCT, MCV, MCH, MCHC,RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR,P-LCC

3-part differentiation of WBC, 21 parameters and 3 histograms

Principles of operation : WBC/RBC/PLT: Electrical Impedance
HGB : Photoelectric colorimetry
Aspiration volume : Whole blood 10μL,Pre-diluted 20μL
Aperture diameter : WBC 100μm,RBC/PLT 68μm
Throughput : 60 samples per hour
Data storage : More than 100,000 sample results with histograms

Dilution ration | Whole blood | Capillary blood
WBC/HGB | 1:232 | 1:400
RBC/PLT | 1:40000 | 1:45000

Display : 10.4 inches color LCD
Control and calibration : L-J,X,X-R,X-B four control modes, manual and auto-Calibration
Dimension : 530mm(L) x 330mm(W)  x 460mm (H)
Power requirement : AC 100V – 240V,  50/60Hz,  180W
Input/Output :

  • Support RS232, standard network port and USB
  • Shortcut keys with standard mouse and keyboard
  • Internal 57mm thermal printer,

Parallel Port and USB for external printer